Dental Implant

How does intraoral photogrammetry actually work?

The Shining 3D Elite intraoral scanner in a hand, and a diagram of its coded scan bodies around an edentulous arch

Firstly, how does a normal intraoral scanner work?

Diagram of structured light scanning: a projector shines straight stripes onto a ridge and a camera at a different angle sees the stripes bend where the surface rises
Straight stripes look bent where the surface rises.

The scanner projects a known pattern of light, such as straight lines, onto the surface. A camera looks at it from a slightly different angle. Where the surface curves the lines look bent, and the amount of bend tells the software how far away that point is.

Summary: A normal scanner measures the shape of whatever it sees: teeth, gum and scan bodies.

What is the weakness of full arch intraoral scanning?

Diagram of a full-arch scan stitched from overlapping frames, with the stitched path moving further from the real arch towards the end of the span
Dashed line: the real arch. Red line: where the stitched frames ended up.

Each 3D snapshot is joined to the one before it by matching the surface they share. Teeth give the software plenty of edges to match. A healed edentulous ridge is smooth, wet and moves, so there is less to lock onto, and each small joining error carries into the next frame.

Summary: The longer the span and the smoother the tissue, the more a scan can cross stitch.

How does a normal scanner find the implant?

Diagram of a best fit: a library file of the scan body is lined up with the scanned mesh, and the fitted axis sits slightly off the true axis
The library file is fitted to the scanned mesh.

It scans the surface of the scan body and builds a mesh. The software then lines a perfect digital copy of that scan body up with the mesh, which is a best fit match. Any noise or cross stitching in the mesh moves where the copy lands, and that becomes the implant position.

Summary: With a normal scan, the implant position depends on the quality of the scanned mesh.

What is photogrammetry?

Diagram of two cameras a known distance apart, each with a line of sight to the same dot; the dot sits where the two lines cross
Where the two lines cross is the dot.

Photogrammetry means measuring from photographs. Using two cameras, a known distance apart, that both see the same point, the software can work out exactly where that point sits in space. Your two eyes do the same thing when you judge depth.

Summary: Two views of the same dot tell you exactly where that dot is.

A measuring method more than 150 years old

Four panels showing photogrammetry in other fields: mapping from the air, the Webb telescope mirrors, A380 wing fittings and motion capture suits

Aimé Laussedat was mapping land from photographs in France in the 1850s, and Albrecht Meydenbauer named the method in 1867.1,2 It has since been used to map countries from the air, check the mirrors of the James Webb Space Telescope,3 measure Airbus A380 wing fittings and track the markers on Andy Serkis’s suit when he played Gollum.

Think of a ceph tracing: you mark the landmarks and measure between them.

Summary: The Shining Elite uses the same measuring method, from inside the mouth.

How is intraoral photogrammetry different from extraoral photogrammetry?

Two arch diagrams: on the left an extraoral camera outside the mouth, where a post in front blocks the view of one further back; on the right the intraoral scanner tip sees two coded scan bodies in one view
Camera outside: a post in front can block the view of one further back. Camera inside: two scan bodies in one view, seen from close up.

Extraoral and intraoral photogrammetry use the same principle. Extraoral systems use a larger camera held outside the mouth, about 15 to 30 cm away. The Shining Elite uses small cameras in the scanner tip. Inside the mouth it can capture two scan bodies in one frame and see past ones that would block a camera outside, which helps with posterior, pterygoid and zygomatic implants.

Summary: Same maths. The camera is just closer to the scan bodies.

The Shining 3D Elite functions in two main modes

The Shining 3D Elite in a hand, beside a diagram of its scanner tip with the projector at the top, two cameras below it and a light under each camera
The Shining 3D Elite and its scanner tip.

Normal scan: The projector shines its light pattern so the cameras can measure the shape of the teeth and gum.

Photogrammetry: The scanner lights up the scan bodies, and the two black and white cameras look only for the dots.

The photogrammetry tip is also larger: 19 × 14 mm against 16 × 12 mm.

Summary: Same scanner, two completely different measurements and modes depending on what is being acquired.

What makes the coded scan bodies different?

Two flat black coded scan bodies drawn in 3D, each with white dots on the top and sides; red lines measure the distances between dots on each and they differ
Red lines: the distances between dots measured on each, and they differ. Not drawn to scale or position exactly.

Every scan body is the same shape with a different dot pattern. The coded scan bodies screw onto the multi-unit abutments. Every one has the same shape: flat and black, with white retroreflective dots on the top, the sides and the screw end. Only the pattern of the dots changes, so the distances between the dots are different on each one.

Seat them facing the centre of the arch, close together but not touching.

Summary: The scanner doesn’t need to measure the shape of the scan body. It only needs to acquire the dots, and capture the distance relationship between the dots.

How does the scanner acquire only the dots and nothing else?

A black and white clinical photo of coded scan bodies seated in an open flap, beside the same view as the cameras see it: white dots on pure black; below, sensor gain turned right down and light turned right up
What you see, and what the cameras are set up to see.

The camera is turned down and the light turned up. In photogrammetry mode the image sensor’s gain, the same idea as ISO on a camera, is turned right down so the picture is almost black, and the light is turned right up. The dots send much more light back towards the cameras than surrounding surfaces do. The system adjusts illumination and camera settings so the dots remain clear while the background stays dark.

Summary: The surrounding features like gingiva never show up in the image, so there is nothing to filter out.

Why do the dots light up?

Three diagrams of light hitting a surface: a mirror bounces it off at an angle, gum scatters it weakly everywhere, and a retroreflective dot sends it straight back to the light source

The dots are retroreflective, like the tape on a safety vest. Tiny glass beads or prisms send light straight back to where it came from. The Elite’s lights sit right beside its cameras, so the dots shine back at them while the gum around them stays dim. A plain white dot wouldn’t work: the sensor gain would have to come back up, and the camera would start picking up colour and texture.

Retroreflective targets send back 100 to 1,000 times more light than a plain white target.4

Summary: The dots return the light to the scanner. The gum sends very little back.

How does it find the exact centre of each dot?

Two pixel grids of the same dot. On the left the brightest single pixel is marked. On the right the weighted balance point of all the lit pixels is marked between pixels, to up to 1/50th of a pixel
The pixel answer and the centroid answer. The numbers show how much of each pixel the dot covers.

A camera sensor is a grid of tiny squares called pixels. Each dot lands across a few of them, bright in the middle and dimmer at the edges. The software weighs every pixel by how bright it is and finds the balance point, which usually falls between pixels, to about 1/50 of a pixel.4,5

That is like finding the middle of a 1 m floor tile to within 2 cm.

Summary: Each dot becomes one precise centre point, finer than the pixels themselves.

How do two cameras turn the dots into 3D points?

Diagram of two coded scan bodies in one frame, with lines of sight from each dot to camera 1 and camera 2
One frame, two scan bodies.

Each camera finds the centre of every dot in its own image. The software matches each dot between the two images and, because it knows how far apart the cameras are, works out where the dot sits in 3D. One frame can hold the dots of two scan bodies, which links neighbouring implants directly.

Summary: Each frame gives a 3D map of dot centres.

How does the scanner know which scan body it is looking at?

Three steps: dot centres measured as crosses, the pattern recognised like a barcode as scan body number 4, and the library file placed so its built-in dots land on the measured centres

The dot pattern works like a barcode. Each coded scan body has its own 3D pattern of dots. Once the scanner has seen enough of them, it knows which scan body it is looking at, the same way a checkout scanner reads a barcode. The software holds a perfect digital file of every coded scan body with the centre of each dot built in, and places it so those dots sit exactly on the measured centres.

Summary: There is no scanned surface to match, so surface noise doesn’t come into it.

What does the lab receive?

A straight multi-unit abutment with a dashed coded scan body above it, marking the centre of the connection, the height of the platform and the angle of the abutment
Scan body: only its dots are measured, then the file is discarded.

The lab gets the position of the multi-unit abutment. Once the scan body file is placed, the software works back to the multi-unit abutment underneath: the centre of the connection, its height and its angle. That is what gets exported for the design. The scan body file is discarded, and the implant system you picked isn’t built into the measurement at all.

Summary: The lab designs on the abutment position, measured through the dots.

Why do saliva and blood stop the dots reading?

Two diagrams of a glass bead: dry, it focuses the light and sends it back; wet, the water drop bends the light and it scatters away
Dry: the bead focuses the light and sends it back. Wet: the drop bends the light and it scatters away.

Each glass bead in a retroreflector is a tiny lens that focuses light onto a mirror behind it, and it only works with air around it. Cover it with water or saliva and the lens stops focusing, so the light scatters instead of coming back. It’s why road lines almost disappear on a wet night: in one road study, beaded line paint lost about 99% of its reflection under simulated rain.6

Summary: Dry the scan bodies before you scan, and re-dry any dot that won’t read.

Diagram of a bead covered in blood soaking up the light, beside a chairside checklist for photogrammetry scans
The light is soaked up.

Blood floods the beads the same way water does. It also has haemoglobin, which soaks up much of the light that reaches it, so very little gets back to the cameras and the dot looks dark.7,8

Chairside checklist

  • Rinse and air-dry the scan bodies before the photogrammetry scan.
  • Keep blood and saliva off the top surfaces while you scan.
  • If one scan body won’t appear, clean just that one and scan it again.
  • Seat every scan body fully. The dots can’t correct one that isn’t seated.

Summary: Photogrammetry takes away the stitching, but you still need clean, dry, fully seated scan bodies.

When can you move on to the next scan body?

Two frames from the Shining 3D Elite user guide: dots still building in orange and green, then the scan bodies recognised as solid models on screen
Screen images: Shining 3D Aoralscan Elite user guide.9

Once the scan body appears, you can move on. While you scan, the dots show orange, then light green, then dark green. Once the scan body model appears on screen, its position has been captured. Acquiring every last dot to dark green improves the result by less than 10 microns, so it may not be clinically significant in some scenarios.

Summary: When the scan body appears, the data acquired is generally sufficient enough. Scanning until all the dots are green may only help a bit more.

How accurate is intraoral photogrammetry?

Trueness is how close a result is to the real position. Precision is how close repeated scans are to each other. Three studies have tested the Elite on full-arch implant positions, one on a lab model and two in patients.10,11,12

Bar chart of linear trueness in a lab study of five photogrammetry systems, lower is better: PIC 17 µm, iCam4D 18 µm, Elite 27 µm, Grammee 28 µm, OxoFit 30 µm; beside two targets showing true and precise versus precise but not true
Linear trueness in the lab study. Lower is better.

In a lab study of five systems on one model, the Elite measured 27 µm, inside the extraoral range.10 On distance, the biggest gap between the Elite and any of the extraoral cameras was about 10 µm (27 against 17 µm for PIC). That is well under the recommended misfit thresholds for passive fit, so clinically it is insignificant.13 There is also more error to manage further down the workflow, in milling for example. On angle the Elite sat mid-pack at 0.27°, ahead of PIC. In the lab, intraoral photogrammetry landed inside the range of the extraoral cameras, at a far lesser cost.

In two clinical studies, one in the maxilla and one in the mandible, it had the best angular trueness of every method tested.11,12

Study Setting Elite, linear Elite, angular How it compared
Lab10 Cast, 6 implants 27 µm 0.27° Inside the range of four extraoral cameras (17 to 30 µm)
Maxilla11 1 patient, 4 implants 33 µm 0.12° Best angle of the four techniques
Mandible12 1 patient, 4 implants 37 µm 0.44° No difference on distance; best angle

Summary: Intraoral photogrammetry landed in the same range as the extraoral cameras in multiple studies.

How it works, in five steps

  1. Sensor gain down, light up: only the dots show.
  2. The centre of each dot is found to about 1/50 of a pixel.
  3. Two cameras place every dot centre in 3D.
  4. The dot pattern identifies the scan body and its file is placed.
  5. The lab receives the multi-unit abutment positions.
Left: the Shining 3D Elite with its coded scan bodies and kit. Right: Dr Jonathan Ng at the Shining 3D headquarters in Hangzhou, China
Left: the Elite, its coded scan bodies and the kit. Right: me at the Shining 3D headquarters in Hangzhou, China.

One scanner for the implants, the soft tissue and the bite. The photogrammetry pass gives the implant positions, read from the dots. The normal scan gives the gingiva, hard tissue, and bite, merged in the same software.

Frequently asked questions

Does intraoral photogrammetry replace the normal intraoral scan?

No. The photogrammetry pass reads only the implant positions from the dots. The gingiva, hard tissue and bite still come from a normal scan, and the Elite merges the two in the same software.

Is intraoral photogrammetry as accurate as extraoral photogrammetry?

In a lab study of five systems the Elite measured 27 µm linear trueness, inside the 17 to 30 µm range of the four extraoral cameras. The biggest gap was about 10 µm, well under the recommended misfit thresholds for passive fit.

Why won’t a scan body read?

Usually it is wet, bloody or not fully seated. Water and saliva stop the beads focusing the light, and blood also soaks the light up. Clean and dry just that scan body and scan it again.

Do all the dots need to turn dark green?

Not always. When the scan body appears, the data acquired is generally sufficient enough. Acquiring every last dot to dark green improves the result by less than 10 microns, so it may not be clinically significant in some scenarios.

About the author

Dr Jonathan Ng, BDSc (Hons), MSc Advanced Dental Implantology, Fellow ICOI, is an implant dentist in Brisbane and the Federal President of the Australian Osseointegration Society. He has placed more than 2,000 implants since 2017 and teaches full-arch reconstruction to dentists across Australia and Asia through Source Education.

References

  1. Grimm A. The origin of the term photogrammetry. In: Fritsch D, ed. Photogrammetric Week ’07. Heidelberg: Wichmann; 2007:53-60.
  2. Polidori L. On Laussedat’s contribution to the emergence of photogrammetry. Int Arch Photogramm Remote Sens Spatial Inf Sci. 2020;XLIII-B2-2020:893-899. doi:10.5194/isprs-archives-XLIII-B2-2020-893-2020
  3. ESA/NASA. James Webb Space Telescope inside the chamber, image 60355 (credit NASA/Chris Gunn). 15 September 2017.
  4. Geodetic Systems Inc. Basics of Photogrammetry. 2017. https://www.geodetic.com/basics-of-photogrammetry/
  5. Luhmann T. Eccentricity in images of circular and spherical targets and its impact to 3D object reconstruction. Int Arch Photogramm Remote Sens Spatial Inf Sci. 2014;XL-5:363-370. doi:10.5194/isprsarchives-XL-5-363-2014
  6. Gibbons RB, Hankey J, Pashaj I. Wet Night Visibility of Pavement Markings. FHWA/VTRC 05-CR3. Virginia Transportation Research Council; 2004.
  7. Bosschaart N, Edelman GJ, Aalders MCG, van Leeuwen TG, Faber DJ. A literature review and novel theoretical approach on the optical properties of whole blood. Lasers Med Sci. 2014;29(2):453-479. doi:10.1007/s10103-013-1446-7
  8. Richert R, Goujat A, Venet L, et al. Intraoral scanner technologies: a review to make a successful impression. J Healthc Eng. 2017;2017:8427595. doi:10.1155/2017/8427595
  9. Shining 3D. Aoralscan Elite user guide and support article: Coded Scan Body Scan. support.shining3ddental.com, accessed 12 November 2026.
  10. Revilla-León M, Gómez-Polo M, Drone M, Barmak AB, Kois JC, Pérez-Barquero JA. Accuracy of complete arch implant scans recorded by using intraoral and extraoral photogrammetry systems. J Prosthet Dent. 2025;134(6):2508-2514. doi:10.1016/j.prosdent.2025.01.041
  11. Revilla-León M, Cascos R, Barmak AB, Drone M, Kois JC, Gómez-Polo M. Accuracy of complete arch implant scans recorded by using intraoral and extraoral photogrammetry systems and a noncalibrated splinting technique: a clinical study. J Prosthet Dent. 2026;135(5):955-962. doi:10.1016/j.prosdent.2025.07.027
  12. Revilla-León M, Cascos R, Barmak AB, Drone M, Kois JC, Gómez-Polo M. Clinical accuracy of complete arch implant scans recorded by using a noncalibrated splinting technique, intraoral photogrammetry, and extraoral photogrammetry with snap-on markers. J Prosthet Dent. 2026;135(1):108.e1-108.e7. doi:10.1016/j.prosdent.2025.09.012
  13. Sallorenzo A, Gómez-Polo M. Comparative study of the accuracy of an implant intraoral scanner and that of a conventional intraoral scanner for complete-arch fixed dental prostheses. J Prosthet Dent. 2022;128:1009-1016.