Full Arch Implant Acquisition Methods: Which Is Most Accurate?

Four full arch implant scanning methods drawn on an edentulous arch: vertical scan bodies, horizontal gauges, intraoral photogrammetry and extraoral photogrammetry

Contents

  1. What are the ways to capture implant position digitally?
  2. What is the difference between precision and trueness?
  3. Can a scanner be precise and still wrong?
  4. How do studies measure linear trueness?
  5. Does “36 µm” mean every implant is within 36 µm?
  6. What is angular trueness, and is there a safe limit?
  7. What was the scan measured against?
  8. Why does a full-arch scan drift across the arch?
  9. Why are vertical scan bodies hard to scan well?
  10. Why link the scan bodies together?
  11. Are all splints and scan aids the same?
  12. How do horizontal scan gauges fix the problem?
  13. Did non-calibrated gauges beat ordinary scan bodies?
  14. How close does intraoral photogrammetry get to extraoral cameras?
  15. Is tablet or phone photogrammetry ready?
  16. How do the methods compare?
  17. What do I use, and which system is best?
  18. Frequently asked questions

What are the ways to capture implant position digitally?

Every method chases the same thing: rigid landmarks the software can lock onto. Why the microns matter is in my post “Misfit in implant supported prosthesis: facts and myths”.

Six ways to capture implant positions, each drawn on an edentulous arch: 1 vertical scan bodies, non-splinting; 2 splint or printed aid, non-calibrated splinting; 3 horizontal gauges, not calibrated, non-calibrated splinting; 4 horizontal gauges, calibrated, calibrated scan bodies; 5 intraoral photogrammetry, calibrated markers; 6 extraoral photogrammetry, camera or tablet

Takeaways: The drawback of digital acquisition of implants is cross stitching, soft tissue and lack of hard tissue references/landmarks.

What is the difference between precision and trueness?

Four targets: precise and true, which is accurate, with the scans clustered on the model; precise, not true, clustered away from it; true on average, not precise, scattered around it; and neither

Scan the same model 100 times. Precision is how close those 100 scans are to each other. Not to the model, just amongst themselves. Trueness is how close those 100 scans are to the model itself. Accuracy needs both.

Summary: Precision means repeatable. Trueness means correct. Accuracy is both together.

Can a scanner be precise and still wrong?

Two targets: precise, not true, where the scans cluster away from the model and give the same wrong answer every time; and true on average, not precise, where the scans scatter but their average lands on the model

If my scanner gives me 100 scans that are all the same, great. But if they’re not the model (or object being scanned), what’s the point? It works the other way too. Trueness is usually reported as an average, so a scanner whose scans scatter all over the place can still average out to the right answer.

Summary: A scanner can repeat the same wrong answer every time, and a scattered one can still look true on average. Always ask for both numbers.

How do studies measure linear trueness?

Six implants numbered around an edentulous arch with a line between every pair; the five distances from implant 1 are red: 15.04 mm to implant 2, 23.41 mm to implant 3, 46.53 mm to implant 4, 47.28 mm to implant 5 and 49.00 mm to implant 6
Red: the five distances from implant 1.

Start with a master model where the true distance between every pair of implants is known. Scan it, say, 20 times, and measure the same distances on every scan. Six implants give 15 distances. Compare each with the master and average the differences. That average is the linear trueness.

Summary: Each straight line between two implants is a Euclidean distance. Linear trueness is the average error across all of them.

Does “36 µm” mean every implant is within 36 µm?

The same five distances from implant 1 as above, now scanned:

Distance On the model On the scan Error
1 to 2 15.04 mm 15.06 mm 20 µm
1 to 3 23.41 mm 23.38 mm 30 µm
1 to 4 46.53 mm 46.57 mm 40 µm
1 to 5 47.28 mm 47.24 mm 40 µm
1 to 6 49.00 mm 49.05 mm 50 µm
Add the errors 180 µm
÷ 5 distances = 36 µm

Illustrative example. A real study averages all 15 distances.

It’s an average. It is not the maximum deviation, not cumulative and not a limit. Some pairs will be better than the mean and some worse. It also can’t tell you which implant is out, or in which direction.

Summary: The 1 to 6 distance was out by 50 µm. A mean linear trueness is never the worst error you will see.

What is angular trueness, and is there a safe limit?

A 14.8 mm implant pivoting at its platform: a tilt of 0.194 degrees moves the apex 50 µm sideways; for two implants, 2 × 0.194 = 0.39 degrees, rounded to 0.4 degrees; the angle is exaggerated to be visible

Angular trueness is the same idea for the implant axes: how far the scanned axes tilt from the true ones, in degrees. Most papers call 0.4 degrees acceptable. That number is mainly a calculation, rather than a clinical finding: Andriessen 20141 assumed an implant can shift 50 µm sideways in bone and worked out the tilt that would cause. The best systems now average about 0.2 degrees.2,3 Plain scan bodies average about 0.5.4

Summary: The 0.4 degrees most papers use is a calculation. The best systems now average about 0.2 degrees, plain scan bodies about 0.5.

What was the scan measured against?

Measured against Accuracy
Coordinate measuring machine
a stylus touches each implant
about 1 to 4 µm5,6
Laboratory or industrial scanner
the cast is scanned on the bench
3 to 5 µm7,8
Extraoral photogrammetry
measured across studies; used as the reference in patients
10 to 77 µm9
Intraoral scanner, plain scan bodies
for comparison, across 100 studies
41 to 557 µm4

When a rep tells you their scanner is 20 microns, ask the question deeper: what did you actually measure that against? You can’t fit a lab scanner in a mouth, so clinical studies have to use photogrammetry as their reference, which is less certain.

Summary: A number is only as good as its reference. Against a CMM or a lab scanner, trust it more. Against another intraoral scanner, be careful.

Why does a full-arch scan drift across the arch?

A full-arch intraoral scan of a mandible with six scan bodies numbered 1 to 6, lines from implant 1 to the others, and mean errors labelled along the arch: 29, 12, 29, 82, 118 and 149 µm
Repeated TRIOS 3 scans of one mandible: the mean error in the distance from implant 1 to each implant in turn (Orejas-Perez 202210).

A scanner stitches small 3D frames together like a jigsaw. Teeth give it plenty to match; an edentulous ridge, palate and blood give it very little. One wrong stitch is carried into every frame after it, so the error grows towards the far side. Near implant 1 the repeated scans agreed within about 30 µm. Once the scan crossed the midline it rose to 82, 118 and 149 µm, and to 204 µm at the patient’s eighth implant.10

Summary: Error builds up as you sweep, so it is worst at the far end of the arch.

Why are vertical scan bodies hard to scan well?

Three views of an intraoral scanner around a tall vertical scan body: 1 over the top, 2 round one side, 3 round the other

To capture a tall post you go over the top, then round one side, then round the other. Every change of angle adds stitching error. The best scans hold one angle and glide. And the scan body in your hand isn’t calibrated, it is not exactly its digital library file: six commercial brands all showed manufacturing tolerances, mostly in height (Lerner 202111).

Summary: A tall thin post makes you change angle at every implant, and the post itself is not a perfect copy of its library file. Across 100 studies, plain scan bodies averaged 113 µm and 0.48 degrees from the reference (Revilla-León 20264).

Two arches: plain scan bodies with only soft tissue between them, and linked scan bodies with something rigid in every gap

The scanner stitches each new frame onto the last by matching shapes. Between plain scan bodies there is only gum: smooth, with few features, and it moves. With little to match, stitches slip, and every slip carries on across the arch.

Linking the scan bodies means you are scanning something hard and rigid rather than soft tissue, all the way round. Splints, scan aids and horizontal gauges are all versions of this idea (Azevedo 202512).

Splinting can cut the error at the two back implants, where it was worst: 142.7 to 96.6 µm at one, 0.61 to 0.49 degrees at the other (Pozzi 20227).

Summary: Soft tissue gives the scanner nothing solid to stitch to. Linking the scan bodies means you scan hard, rigid geometry instead.

Are all splints and scan aids the same?

Three photos of the same mandibular model: (A) plain scan bodies, (B) Straumann EXACT linked scan bodies, (C) ioConnect horizontal gauges
(A) plain scan bodies (B) Straumann EXACT (C) ioConnect, one model (Blivajs 202613). Photo: Blivajs, Dent J 2026, CC BY 4.0.
Product What the study found
Straumann EXACT (linked scan bodies) 0.19 vs 0.23 mm and 1.3 vs 2.7° against plain scan bodies (Blivajs 202613)
Scan Ladder (textured bars) 18.4 µm, the best of four aids on one model (Lin 202614)
Apollo SmartFlag (PEEK wings) 24.6 µm, second of four (Lin 202614); the most accurate of five (Laureti 202515)
PoLo and M6 (horizontal arms) a plain vertical scan body did better (Laureti 202515)
OptiSplint (frame) 30.9 vs 56 µm without it (Prott 202516)

Summary: No. Some helped a lot; some did no better than a plain scan body. Check the study for your product.

How do horizontal scan gauges fix the problem?

Two horizontal scan gauge systems on an arch: calibrated Nexus IOS gauges that overlap and are scanned once from each side, and non-calibrated ioConnect gauges angled to the centre, where only the hexagons are scanned

Calibrated gauges are each measured on a CMM after they are made, so your kit gets its own library file. Non-calibrated gauges use a standard one.17

Every Nexus study against a plain scan body Nexus Plain
In vitro, 6 implants, 4 scanners (Azevedo 202512) 10 to 40 µm 20 to 95 µm
In vitro, 4 implants, 4 scanners (Laureti 202515) 11 to 40 µm 11 to 31 µm
In vitro, 4 implants; the maker processed the Nexus scans (Lerner 202618) about 14 µm about 30 µm
In vitro, 4 and 6 implants: 12 prosthodontists did the one-screw test on bars milled from each scan (Kerzhner 202519) 10 and 6 passed 0 passed

In the two independent studies, Nexus was clearly better in 3 of 8 scanner and operator pairings in each, and no different in the rest. In patients, all 37 arches passed the one-screw test, with no comparison group (Klein 202320).

Summary: Independent studies found Nexus better than a plain scan body in some set-ups and level in the rest.

Did non-calibrated gauges beat ordinary scan bodies?

Dot chart of deviation in µm with ioConnect gauges against plain scan bodies on five intraoral scanners: Primescan 26 against 78, TRIOS 5 30 against 70, Aoralscan 3 34 against 72, Medit i700 41 against 195, iTero Element 5D 58 against 87
● with ioConnect   ○ plain scan bodies   lower is better

With gauges angled to the palate, every scanner got closer to the reference (Revilla-León 202621). Apollo’s SmartFlag, tested the same way, mainly improved the angles, about 0.28 against 0.30 to 0.46 degrees, and distance did not change significantly (Revilla-León 202522).

In vitro, one maxillary cast, six implants, 10 scans each, against a lab scanner.

Summary: Mostly yes, and the technique mattered more than the scanner. But gauge designs don’t all behave the same.

How close does intraoral photogrammetry get to extraoral cameras?

Bar chart of linear trueness for five photogrammetry systems tested on the same model: PIC (extraoral) 17 µm, ICam4D (extraoral) 18 µm, Elite (intraoral) 27 µm, Grammee (extraoral) 28 µm, OxoFit (extraoral) 30 µm
Data from Revilla-León 20259.

In a lab study of five systems on one model, the Elite measured 27 µm, inside the extraoral range.9 In two clinical studies it had the best angular trueness of every method tested.23,24 In patients, it deviated 30.7 µm against 59.8 µm for plain scan bodies, and 0.78 against 1.4 degrees (Eldabe 202625).

Summary: Within about 10 µm of the best extraoral cameras in the lab.

Is tablet or phone photogrammetry ready?

A tablet held above four coded scan bodies on the implants, photographing their dot patterns
Study App Compared with
T-Marker on iPad, lab (Revilla-León 202626) 88 µm gauges 44 µm
T-Marker on iPad or iPhone, lab (Revilla-León 202627) 39 to 58 µm gauges 42 to 47 µm
PIC app on iPhone, lab, against a CMM (Santamaría-Laorden 202528) 31 µm PIC camera 18 µm
PIC app on iPhone, lab (Omar 202629) 23 µm MicronMapper 22 µm
T-Marker on iPhone, 24 patients (Alghraisi 202630) 50 µm intraoral photogrammetry 34 µm

Across nine studies,26–34 the apps ranged from 17 to 88 µm depending on markers, device and reference, and the T-Marker’s angles beat the gauges’ in all three comparisons.26,27,31

Summary: Promising results have been shown with tablet or phone based photogrammetry, cutting the equipment costs required for photogrammetry by a huge margin.

How do the methods compare?

Average deviation from a lab reference (lower is better)

Method Distance Angle Evidence
Plain vertical scan bodies 113 µm 0.48° review, 100 studies4
Splint or printed aid all within 150 µm review, 12 studies35
Horizontal gauges, not calibrated 26 to 58 µm (ioConnect); 45 to 100 µm (SmartFlag) 0.25 to 0.39° lab, five scanners each21,22
Horizontal gauges, calibrated 10 to 40 µm (surface measure) two independent lab studies12,15
Intraoral photogrammetry 36 µm 0.23° review2
Extraoral photogrammetry 26 µm 0.21° review3
Phone or tablet photogrammetry 17 to 88 µm 0.07 to 0.54° nine studies, mostly lab26–34

Summary: Operator technique is one of the most important factors as well as the quality of your milling process. Don’t hang your hat on just the published figures as acquisition is only the first part of the workflow.

What do I use, and which system is best?

Black and white clinical photo of Shining 3D Elite intraoral photogrammetry scan bodies seated on a lower arch after surgery
Shining 3D Elite intraoral photogrammetry scan bodies, seated after surgery.

For every full arch, I capture the implant positions with intraoral photogrammetry. It performs very close to extraoral photogrammetry, and my results are tight and consistent and my auxiliary team is familiar with an intraoral scanner setup. I also can get to the posterior regions easily.

Summary: Which is best for you is multifactorial: cost against how often you’ll use it, the skill it needs, ease of use, accuracy, your lab’s support, and the one nobody talks about, your lab’s milling process. The most accurate camera can’t fix a cheap mill.

Frequently asked questions

What is the difference between precision and trueness?

Scan the same model 100 times. Precision is how close those 100 scans are to each other. Trueness is how close those 100 scans are to the model itself. Accuracy needs both.

Does a mean linear trueness of 36 µm mean every implant is within 36 µm?

It’s an average. It is not the maximum deviation, not cumulative and not a limit. Some pairs will be better than the mean and some worse. A mean linear trueness is never the worst error you will see.

Is 0.4 degrees a safe limit for angular trueness?

Most papers call 0.4 degrees acceptable. That number is mainly a calculation, rather than a clinical finding. The best systems now average about 0.2 degrees, plain scan bodies about 0.5.

Which full arch implant acquisition method is most accurate?

In the reviews, extraoral photogrammetry averaged 26 µm and 0.21° from a lab reference, intraoral photogrammetry 36 µm and 0.23°, and plain vertical scan bodies 113 µm and 0.48°. Which is best for you is multifactorial: cost against how often you’ll use it, the skill it needs, ease of use, accuracy, your lab’s support, and your lab’s milling process.

About the author

Dr Jonathan Ng, BDSc (Hons), MSc Advanced Dental Implantology, Fellow ICOI, is an implant dentist in Brisbane and the Federal President of the Australian Osseointegration Society. He has placed more than 2,000 implants since 2017 and teaches full-arch reconstruction to dentists across Australia and Asia through Source Education.

References

  1. Andriessen FS, Rijkens DR, van der Meer WJ, Wismeijer DW. Applicability and accuracy of an intraoral scanner for scanning multiple implants in edentulous mandibles: a pilot study. J Prosthet Dent. 2014;111(3):186-194. doi:10.1016/j.prosdent.2013.07.010
  2. Revilla-León M, Ghunaim D, Barmak AB, Afshari FS, Fang Q, Prasad S. Accuracy of complete arch implant scans recorded by using calibrated scanning techniques: a systematic review and meta-analysis. Report of the Committee on Research in Fixed Prosthodontics of the American Academy of Fixed Prosthodontics. J Prosthet Dent. 2026;135(4):e28-e42. doi:10.1016/j.prosdent.2025.11.029
  3. Prasad S, Barmak AB, Revilla-León M. A systematic review and meta-analysis of the accuracy of complete arch implant scanning techniques performed extraorally: reverse impression and extraoral photogrammetry methods. Report of the Committee on Research in Fixed Prosthodontics of the American Academy of Fixed Prosthodontics. J Prosthet Dent. 2026;136(1):e79-e101. doi:10.1016/j.prosdent.2026.01.018
  4. Revilla-León M, Barmak AB, Prasad S. Accuracy of complete arch implant scans using nonsplinting techniques: a systematic review and meta-analysis. Report of the Committee on Research in Fixed Prosthodontics of the American Academy of Fixed Prosthodontics. J Prosthet Dent. 2026;136(1):e102-e138. doi:10.1016/j.prosdent.2026.02.034
  5. Revilla-León M, Att W, Özcan M, Rubenstein J. Comparison of conventional, photogrammetry, and intraoral scanning accuracy of complete-arch implant impression procedures evaluated with a coordinate measuring machine. J Prosthet Dent. 2021;125(3):470-478. doi:10.1016/j.prosdent.2020.03.005
  6. Spagopoulos D, Kaisarlis G, Spagopoulou F, Halazonetis DJ, Güth JF, Papazoglou E. In vitro trueness and precision of intraoral scanners in a four-implant complete-arch model. Dent J (Basel). 2023;11(1):27. doi:10.3390/dj11010027
  7. Pozzi A, Arcuri L, Lio F, Papa A, Nardi A, Londono J. Accuracy of complete-arch digital implant impression with or without scanbody splinting: an in vitro study. J Dent. 2022;119:104072. doi:10.1016/j.jdent.2022.104072
  8. Ma B, Yue X, Sun Y, Peng L, Geng W. Accuracy of photogrammetry, intraoral scanning, and conventional impression techniques for complete-arch implant rehabilitation: an in vitro comparative study. BMC Oral Health. 2021;21(1):636. doi:10.1186/s12903-021-02005-0
  9. Revilla-León M, Gómez-Polo M, Drone M, Barmak AB, Kois JC, Pérez-Barquero JA. Accuracy of complete arch implant scans recorded by using intraoral and extraoral photogrammetry systems. J Prosthet Dent. 2025;134(6):2508-2514. doi:10.1016/j.prosdent.2025.01.041
  10. Orejas-Perez J, Gimenez-Gonzalez B, Ortiz-Collado I, Thuissard IJ, Santamaria-Laorden A. In vivo complete-arch implant digital impressions: comparison of the precision of three optical impression systems. Int J Environ Res Public Health. 2022;19(7):4300. doi:10.3390/ijerph19074300
  11. Lerner H, Nagy K, Luongo F, Luongo G, Admakin O, Mangano F. Tolerances in the production of six different implant scan bodies: a comparative study. Int J Prosthodont. 2021;34(5):591-599. doi:10.11607/ijp.7379
  12. Azevedo L, Laureti A, Marques T, Pitta J, Fehmer V, Pozzi A, et al. Effect of horizontal and vertical intraoral scan bodies on the trueness of complete-arch digital implant impressions: a comparative in vitro study with six implants. Clin Oral Implants Res. 2025;36(9):1136-1145. doi:10.1111/clr.14463
  13. Blivajs I, Smojver I, Reiser Y, Jurić S, Hadžović E, Pelivan I, et al. Linear and angular deviations of complete-arch digital implant impressions using three scan body systems: an in vitro study. Dent J (Basel). 2026;14(10):686. doi:10.3390/dj14100686
  14. Lin TY, Wang TM, Lin LD. Accuracy of full-arch implant scans using different horizontal scan bodies in an angulated model: an in vitro study. J Dent Sci. 2026;21(4):2108-2117. PMCID: PMC13689625
  15. Laureti A, Marques T, Pitta J, Fehmer V, Sailer I, Pozzi A, et al. Influence of horizontal intraoral scan bodies on the trueness of digital impressions for complete-arch prostheses on four implants: an in vitro evaluation. Clin Oral Implants Res. 2025;36(10):1287-1295. doi:10.1111/clr.70001
  16. Prott LS, Atria PJ, Maluf CV, Blatz MB, Conejo J. In vitro comparison between complete arch abutment-level implant impressions with photogrammetry, grammetry, and intraoral scanning. J Prosthet Dent. 2025;134(3):782.e1-782.e7. doi:10.1016/j.prosdent.2025.05.009
  17. Revilla-León M, Gómez-Polo M, Rutkunas V, Ntovas P, Kois JC. Classification of complete-arch implant scanning techniques recorded by using intraoral scanners. J Esthet Restor Dent. 2025;37(1):236-243. doi:10.1111/jerd.13322
  18. Lerner H, Weigl P, Sader R, Klein M, Polack MA, Tuminelli FJ. Accuracy of full-arch implant scan with Nexus IOS scan gauges versus different conventional scan bodies: an in vitro comparative study. J Dent. 2026;164:106254. doi:10.1016/j.jdent.2025.106254
  19. Kerzhner S, Lin HC, Tuminelli F. Passivity of complete arch implant-supported fixed dental prostheses fabricated using a scan gauge system and conventional scan bodies: an in vitro study. J Prosthodont. 2025 Jun 14 [Epub ahead of print]. doi:10.1111/jopr.14079
  20. Klein M, Tuminelli FJ, Sallustio A, Giglio GD, Lerner H, Berg RW, et al. Full-arch restoration with the NEXUS IOS system: a retrospective clinical evaluation of 37 restorations after a one year of follow-up. J Dent. 2023;139:104741. doi:10.1016/j.jdent.2023.104741
  21. Revilla-León M, Cascos R, Barmak AB, Kois JC, Gómez-Polo M. Accuracy of a complete arch noncalibrated splinting implant scanning technique with a palatal orientation recorded by using different intraoral scanners. J Prosthet Dent. 2026;135(1):109-117. doi:10.1016/j.prosdent.2025.02.020
  22. Revilla-León M, Cascos R, Lawand G, Barmak AB, Kois JC, Gómez-Polo M. Accuracy of complete arch nonsplinting and noncalibrated splinting implant scanning techniques recorded by using five intraoral scanners. J Prosthet Dent. 2025;133(6):1581.e1-1581.e10. doi:10.1016/j.prosdent.2025.02.062
  23. Revilla-León M, Cascos R, Barmak AB, Drone M, Kois JC, Gómez-Polo M. Accuracy of complete arch implant scans recorded by using intraoral and extraoral photogrammetry systems and a noncalibrated splinting technique: a clinical study. J Prosthet Dent. 2026;135(5):955-962. doi:10.1016/j.prosdent.2025.07.027
  24. Revilla-León M, Cascos R, Barmak AB, Drone M, Kois JC, Gómez-Polo M. Clinical accuracy of complete arch implant scans recorded by using a noncalibrated splinting technique, intraoral photogrammetry, and extraoral photogrammetry with snap-on markers. J Prosthet Dent. 2026;135(1):108.e1-108.e7. doi:10.1016/j.prosdent.2025.09.012
  25. Eldabe AK, Adel-Khattab D, Botros KH. Accuracy of intraoral photogrammetry in complete arch digital implant scanning: an in vivo prospective comparative study. J Prosthet Dent. 2026;135(2):342-351. doi:10.1016/j.prosdent.2025.03.041
  26. Revilla-León M, Gómez-Polo M, Barmak AB, Kois JC, Pérez-Barquero JA. Accuracy of implant scans recorded using tablet-based photogrammetry and noncalibrated implant scan body systems. J Prosthet Dent. 2026;136(2):570-576. doi:10.1016/j.prosdent.2026.03.034
  27. Revilla-León M, Cascos R, Barmak AB, Kois JC, Gómez-Polo M. Influence of number and location of implants and data acquisition device on the accuracy of an app-based extraoral photogrammetry and noncalibrated implant scan body system. J Prosthet Dent. 2026;136(5):1346-1357. doi:10.1016/j.prosdent.2026.08.016
  28. Santamaría-Laorden A, Marugán AM, Andreu-Vázquez C, Orejas-Pérez J. Complete arch implant capture using a photogrammetry algorithm and smartphone app: an in vitro study. J Dent. 2025;161:105969. doi:10.1016/j.jdent.2025.105969
  29. Omar MS, Lan YH, Nagai T, Yang CC, Lin WS. Comparative trueness of complete-arch implant position acquisition using auxiliary geometric device-aided and noncalibrated scanbody-based intraoral scanning versus smartphone-based and dedicated extraoral photogrammetry. J Prosthodont. 2026 Sep 24 [Epub ahead of print]. doi:10.1111/jopr.70244
  30. Alghraisi MI, Elsyad MA, El-Sawy AA, Dakhil R, Zahran K, El-Sawy MA. Accuracy of complete-arch implant scans obtained using intraoral photogrammetry and a mobile-phone based photogrammetry system (T-Marker): a randomized crossover clinical study. J Dent. 2027;176:107046. doi:10.1016/j.jdent.2026.107046
  31. Revilla-León M, Cascos R, Barmak AB, Kois JC, Drone M, Gómez-Polo M. Clinical evaluation of complete arch implant scanning accuracy recorded using an app-based photogrammetry and noncalibrated scan body system. J Prosthet Dent. 2026;136(2):458-465. doi:10.1016/j.prosdent.2026.05.030
  32. Santamaría-Laorden A, Castillo-Oyagüe R, Martínez-Marugán A, Quirós Sánchez JL, Andreu-Vázquez C, Orejas-Pérez J. Accuracy of a smartphone app and photogrammetry algorithm in complete arch digital implant capture: an in vivo comparative study. J Prosthet Dent. 2026;135(4):777.e1-777.e9. doi:10.1016/j.prosdent.2025.11.020
  33. Martínez-Marugán A, Santamaría-Laorden A, Gómez-Cogolludo P, Baldera Rodríguez AP, Andreu-Vázquez C, Orejas-Pérez J. Complete arch digital implant capture using a smartphone-based photogrammetry app: an in vitro comparative study of the accuracy of different fiducial marker designs. J Prosthet Dent. 2026;136(4):e54-e60. doi:10.1016/j.prosdent.2026.06.024
  34. Gianfreda F, Raffone C, Martelli M, Pieralli S, Bollero P, Caponio VCA. Accuracy of smartphone-based photogrammetry for full-arch implant position transfer: a controlled in vitro comparison with scan body-based intraoral scanning. J Dent. 2026;172:106716. doi:10.1016/j.jdent.2026.106716
  35. Revilla-León M, Ghunaim D, Barmak AB, Afshari FS, Fang Q, Prasad S. A systematic review of the accuracy of complete arch implant scans recorded by using noncalibrated splinting or implant scan body techniques. Report of the Committee on Research in Fixed Prosthodontics of the American Academy of Fixed Prosthodontics. J Prosthet Dent. 2026;135(6):1058-1071. doi:10.1016/j.prosdent.2025.11.018